Xyp commands, Calibration and verification – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 91

background image

x

MAP Technology

Software Tour

PN 89-00002-00-070 Rev. A

7 - 27

View Failed Self-

Diagnostic Details

‹

To view details regarding which of the self-diagnostics tests

passed or failed:

1. Select the Diagnostics tab and view the message log. At least one

error from a failed self-diagnostics test appears with a yellow
background.

2. Double click the yellow row for which you want to see a more

detailed description. An Errors dialog box appears showing a list
of passed and failed self-diagnostic tests.

3. Click OK to close the dialog box. A dialog box appears

indicating which of the self-diagnostic tests passed or failed.
Click OK to close this dialog box.

See the Troubleshooting chapter in this manual if the self-diagnostic
routine reports a failure.

XYP Commands

XYP commands let you eject or retract the plate holder on the XYP
instrument, command the Sample Probe up or down, and run self
diagnostics.

Calibration and
Verification

You can run calibration and verification commands from the
Maintenance tab on the main screen.

Set Well Locations
for Calibration and

Controls

‹

To set the locations on the microtiter plate for calibration and

control solutions:

1. Click the drop down arrow at the end of the entry cell. A list of

well locations appears below the arrow.

2. Click the well location on the microtiter plate image.

Enter System

Calibration

Targets

‹

To enter new and changed system calibration targets:

Refer to the Certificates of Quality included with the xMAP
calibrators to find the reagent.