x
MAP Technology
4 - 10
PN 89-00002-00-070 Rev. A
heater block in the bracket inside the left access door of the Luminex
100 analyzer.
To install the XYP instrument heater block:
1. Click the Eject/Retract button to eject the plate holder.
2. Insert the XYP instrument heater block on the plate holder.
Figure 23. The XYP Instrument Heater Block
3. Click the Eject/Retract button to retract the plate holder.
Adjust the Sample
Probe Vertical
Height
You need to adjust the vertical height of the Sample Probe each time
the type or style of microtiter plate is changed.
Figure 24. Thumb Wheel and Height Adjustment Locking Screw
2
1
1. Thumb Wheel
2. Height Adjustment Locking Screw