Luminex 100 IS User Manual, Version 2.1 User Manual
Page 35
x
MAP Technology
Installation
PN 89-00002-00-070 Rev. A
4 - 7
3. Remove the Cheminert fitting by turning the connector
counterclockwise. If the tubing is difficult to remove, push up
gently on the sampling probe.
Figure 19. Cheminert Fitting
4. Remove the sampling probe by grasping the probe and gently
pushing up.
5. Insert the XYP instrument sampling probe and replace the tubing
connector. Tighten the connector by turning the connector
clockwise, being careful that the threads are correctly aligned.
Hand tighten only.
1. Cheminert Fitting
2. Sample Probe
1
2
Caution:
The XYP instrument sampling probe should slide up
easily while removing it from the Sample Arm. If you feel
resistance, do not force the probe up. Contact Customer Support.