beautypg.com

Luminex 100 IS User Manual, Version 2.1 User Manual

Page 111

background image

x

MAP Technology

Software Tour

PN 89-00002-00-070 Rev. A

7 - 47

plate appears on the screen to make room for additional samples. The
new plate appears to the immediate right of the existing plate image
on the screen with a dark line between the adjacent columns of the
two plates.

Figure 78. Processing Multiple Plates

The samples on the next plate use the same template that the samples
in the first plate use.

During acquisition for the batch, when all of the wells on the first
microtiter plate have been tested, the system pauses and prompts you
to insert the next plate.

‹

To process multiple plates during a batch:

1. When you create the batch, enter all the samples you need. New

plates appear automatically as needed.
If you do not know how to create a batch, see “Create a New
Batch” on page 7-39. Click Finish.

Note:

You may need to scroll to the

right of the screen to see the
additional plate(s).