Sample high/low, Failed standard in batch, Sample timed out – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 158
![background image](/manuals/336970/158/background.png)
Luminex 100 IS User Manual Version 2.1
x
MAP Technology
9 - 8
PN 89-00002-00-070 Rev. A
•
Ensure that the correct control lot number and target values were
selected and entered correctly.
•
Verify that the system is functioning properly by running the
system controls.
•
If the system controls pass, contact the kit manufacturer.
Sample High/Low
This message occurs if the sample results do not fall within the
highest and lowest standards.
•
Ensure that you selected the correct template.
•
Ensure that the correct standard lot number and target values
were selected and entered correctly.
•
Verify that the system is functioning properly by running the
system controls.
•
If the sample is noted as High, dilute the sample following the
kit manufacturer’s suggestions.
•
If the system controls pass, contact the kit manufacturer.
Failed Standard in Batch
This message occurs when one or more of the assay standards are not
within the expected range.
•
Ensure that you selected the correct template.
•
Ensure that the correct assay standard lot number and target
values were selected and entered correctly.
•
Verify that the system is functioning properly by running system
controls.
•
If the system controls pass, contact the kit manufacturer.
Sample Timed Out
•
Ensure that the samples were properly mixed before they were
dispensed into the wells.
•
Ensure that you loaded the sample to the correct well on the
plate.
•
Check if the sample volume added to the well is correct.
•
Remove and sonicate or flush the XYP sampling probe to
remove possible clogs.
•
Check if the XYP sampling probe is properly aligned.
•
Complete the following analyzer functions from the Maintenance
tab: Drain, Backflush, and Alcohol Flush.
•
Run the system controls.
•
Check to see if the delta cal temperature < 3.0 degrees.
Recalibrate if greater than 3.0 degrees and retry.
•