Annually, Sheath filter – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 144
100 IS User Manual Version 2.1
x
MAP Technology
8 - 8
PN 89-00002-00-070 Rev. A
Annually
Sheath Filter
To change the sheath filter within the analyzer:
1. Disconnect the Luminex 100 analyzer from ac power by
turning off the power switch on the rear of the Luminex 100
analyzer, then unplugging the instrument power cord from
the wall source.
2. Disconnect the sheath fluid bottle before changing the filter.
3. Open the left door of the Luminex 100 analyzer and disconnect
the filter by pushing down on the metal clamps on each
connection.
Figure 105. Sheath Filter
4. Connect the new sheath filter, matching up the color-coded
fittings. The arrow on the sheath filter should be pointing up.
5. Reconnect the sheath fluid bottle.
6. Prime twice.
7. Close the left door on the Luminex 100 analyzer.
8. Plug in and power on the Luminex 100 analyzer.
Sheath filter