Accuracy and precision, Sensitivity, Capacity – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 46
Luminex 100 IS User Manual Version 2.1
x
MAP Technology
5 - 4
PN 89-00002-00-070 Rev. A
Accuracy and Precision
•
Sample uptake volume ± 5%
•
Classification of microspheres > 80%
•
Misclassification of microspheres < 0.5%
•
Temperature control 0°C to +3°C of target
•
Internal sample carry over < 0.9%
•
Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity values
Sensitivity
•
Detect 1000 fluorochromes phycoerythrin (PE) per microsphere
•
Reporter channel dynamic range: 3.5 decades
Capacity
The specifications below reflect minimum capacity values:
•
10 GB hard drive
•
Store data for up to 240,000 test results
•
1.4 MB 3½” floppy disk
•
100 MB read/write CD ROM
•
Analyze multiple 96-well plates per batch
•
Analyze multiple assay templates per plate
•
Distinguish a minimum of 1 to a maximum of 100 unique xMAP
microsphere sets in a single sample
•
Detect and distinguish surface reporter fluorescence emissions at
575 nm on the surface of 1-100 unique xMAP microspheres sets
in a single sample
•
Reporter channel range: 3.5 decades of detection
•
Sample core: 15 - 20 µm core at 1 µL/sec sample inject rate
•
Maintain samples at a constant temperature from 35°C to 55°C
(95°F to 131°F)
•
Automatic sampling from a 96-well plate
•
Start sampling from any well position
•
Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills
•
Microtiter plates with 96 wells must be compatible with the XYP
instrument plate holder. The following microtiter plate types are
compatible with the XYP instrument plate holder: flat bottom,
conical, round, filter bottom, half plates [overall height no more
than 0.75” (19 mm)], any color
•
Microtiter plates with 96 wells must be compatible with XYP
instrument heater block temperature from 35°C to 55°C (95°F to
131°F) when performing heated assays and using the heater
block.