5 - 13

Air, Waste, and

Sheath Fluid

Connectors

Located on the side of the instrument, these connectors couple
directly to the sheath and waste fluid containers. The air connector is
green, the sheath fluid connector is blue, and the waste fluid
connector is orange.

1. Sheath Connector (Blue)
2. Air Connector (green)
3. Waste Connector (orange)

Figure 32. Air, Waste, and Sheath Fluid Connectors

If your system uses the Luminex Sheath Delivery system, see
“Luminex Sheath Delivery System” in the appendix for more
information about the Sheath Delivery system and its installation.

Sheath Fluid

Container (not

shown)

For proper operation, this container must be placed at the same
level as the base of the XYP instrument.
The sheath fluid container
should not be placed on top of the Luminex 100 analyzer. The sheath
fluid level should be below the air inlet connection and above the
sheath outlet connection.

Waste Fluid

Container (not

shown)

This receives waste from the system. Waste levels must be
manually monitored
—do not allow the waste container to
overflow! Empty the waste container each time the sheath fluid
container is filled. The waste container should not be placed on top
of the instrument.

1

3

2

Note:

Sheath fluid levels

must be monitored
manually. Check the
sheath fluid level before
starting a run or
procedure.

Warning:

If biological samples have been tested with the system,

use your standard laboratory safety practices.