Tabs – Luminex 100 IS User Manual, Version 2.1 User Manual
Page 73
x
MAP Technology
Software Tour
PN 89-00002-00-070 Rev. A
7 - 9
Tabs
The main screen contains four tabs: Home, Run Batch, Maintenance,
and Diagnostics. These tabs contain different features and commands
performed during various stages of the sample acquisition process.
Paused
This term appears when the
software stops processing the list
of commands. A “resume”
function becomes available to
change the state back to
“processing.”
Red
Disconnected
This term appears when the
software has not yet attempted to
connect or fails to connect to the
instrument. To connect the
instrument, power on the
analyzer, then the XY Platform
and click Connect on the toolbar.
Locked Out
This term appears when another
application currently has control
of the instrument. The software
locks out as long as the other
application runs. To remove the
locked out status, close the other
application or wait until the
application completes.
Table 2. Status Bar Communication Message Details
Status Bar
Color
Message
Indicates