OF-500 OptiFiber Certifying OTDR
Users Manual

70

Loss/Length Testing in Loopback Mode (cont.)

7

Optional: From the TEST SETUP screen, you can enter
reference test cord lengths to meet TSB-140 reporting
requirements:

Use

L

to highlight a reference test cord number, press

t

; then enter a length. Press

N

when you are done.

8

Press

C

OK to leave the TEST SETUP screen.

W

Caution

If you disconnect the reference test cord from the
tester's output, you must set the reference again
to ensure valid measurements.

9

Clean the connectors on the cabling to be tested; then
connect the reference test cords and the short patch
cord to the cabling, as shown in Figure 29.

10

Press

T

to start the loss/length test.

11

To save the results, press

s

, select or create a fiber

ID; then press

s

.