Cleaning the loss/length connectors, Cleaning the output connector, Cleaning the input connector – Atec Fluke-OF-500 User Manual
Page 64
OF-500 OptiFiber Certifying OTDR
Users Manual
54
Cleaning the Loss/Length Connectors
Always clean and inspect fiber connectors before making
connections. Use 99 %-pure isopropyl alcohol and optical-
grade wipes or swabs to clean the connectors as follows:
Cleaning the OUTPUT Connector
1
Touch the tip of a 2.5 mm lint-free foam swab in
alcohol; then touch the swab to an alcohol-soaked
cleaning pad.
2
Push the swab into the adapter; twist it around 3 to 5
times against the endface, then remove and dispose
of the swab.
3
Dry the endface with a dry swab by twisting it around
in the adapter 3 to 5 times.
4
Inspect the connector regularly with a fiber
microscope, such as the Fluke Networks
FiberInspector Video Microscope.
Cleaning the INPUT Connector
Note
Typically, the input connector requires cleaning
only if it has been touched.
1
Remove the connector adapter to expose the
photodiode lens (see Figure 21).
2
Use the method described under "Cleaning the
Output Connector" above to dampen a swab with
alcohol.
3
Twist the damp swab around against the lens 3 to 5
times; then twist a dry swab around against the lens 3
to 5 times.