Loss/Length Testing in Smart Remote Mode (cont.)

11

Clean the connectors on the cabling to be tested; then

connect the reference test cords and the short patch
cords to the cabling, as shown in Figure 26.

12

Press

T

to start the loss/length test.

If Open or Unknown appears as the status, try the
following:

Verify that all connections are good.

For an OptiFiber smart remote verify that the
remote tester is set to Remote on its Loss/Length
tab in Setup.

Verify that the remote tester is still active. For an
OptiFiber smart remote you may need to press

A

Start on the remote tester to reactivate the tester.

Try different connections to the cabling until the
test continues. See “Using FindFiber in Smart
Remote Mode” in the online help or Technical
Reference Handbook for details on FindFiber
messages.

Use a visible light source to verify fiber continuity.

13

To save the results, press

s

, select or create a fiber

ID for the INPUT fiber; then press

s

.

14

Select or create a fiber ID for the OUTPUT fiber; then

press

s.