beautypg.com

Atec Fluke-OF-500 User Manual

Page 121

background image

Specifications

111

Power Meter Specifications (cont.)

Power measurement uncertainty (accuracy)

±

0.25 dB

1

Measurement linearity
(18

°

C to 28

°

C constant temperature)

1300/1310 nm and 1550 nm:

±

0.1 dB

2

850 nm:

±

0.15 dB

3

Auto wavelength detect

On OFTM-573x modules when used with a SimpliFiber source

Re-calibration period

1 year

Display update rate

1 reading per second

1.

Under the following conditions:

Power level: -20 dBm, continuous wave

At 850 nm: 62.5/125

µ

m fiber with 0.275 NA

At 1310 nm and 1550 nm: 9 /125

µ

m

Ambient temperature: 23

o

C

±

5

o

C

2.

Linearity for 1300 nm, 1310 nm, and 1550 nm:

Between 0 dBm and -55 dBm:

±

0.1 dB

<

-55 dBm:

±

0.2 dB

3.

Linearity for 850 dBm:

Between -15 dBm and -52 dBm:

±

0.15 dB typical performance

Between 0 dBm and -15 dBm:

±

0.5 dB typical performance