OF-500 OptiFiber Certifying OTDR
Users Manual

62

Loss/Length Testing in Smart Remote Mode (cont.)

6

Clean the tester's and source's OUTPUT connectors and
the reference test cord connectors.

7

On the main tester: press

F

; then select Set

Loss/Length Reference.

8

Connect the reference test cords as shown on the screen
and in Figure 25; then press

t

.

9

Optional: From the TEST SETUP screen, you can enter
reference test cord lengths to meet TSB-140 reporting
requirements:

Use

L

to highlight a reference test cord number, press

t

; then enter a length. Press

N

when you are done.

10

Press

C

OK to leave the TEST SETUP screen.

W

Caution

If you disconnect the reference test cords from the
tester's or remote's output, you must set the
reference again to ensure valid measurements.

-continued-