beautypg.com

Atec Agilent-E8364B User Manual

Page 94

background image

90

Table 30. P11644A Calibration Kit

Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)

Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch

Applies to the, E836xB/C analyzers, P11644A (WR-62) calibration kit, 85132F flexible test port cable set with the P281B and P281C launch
sets, and a full 2-port calibration. Also applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

12.4 to

18 GHz

Directivity

50

Source Match

50

Load Match

50

Reflection Tracking

±0.000

+0.02/°C

Transmission Tracking

±0.016

+0.02/°C

This manual is related to the following products: