9

E8363B/C AND E8364B/C Corrected System Performance with 2.4mm Connectors

⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯

Table 3. 85056A Calibration Kit

Standard Configuration and Standard Power Range

(E8363B/C AND E8364B/C)

Applies to the E8363B/C AND E8364B/C analyzers, 85056A (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

45 MHz to

2 GHz

2 to

20 GHz

20 to

40 GHz

40 to

50 GHz

Directivity

42

42

38

36

Source Match

41

38

33

31

Load Match

42

42

37

35

Reflection Tracking

±0.001

+0.02/°C

±0.008

+0.02/°C

±0.020

+0.02/°C

±0.027

+0.03/°C

Transmission Tracking

±0.010

+0.02/°C

±0.049

+0.02/°C

±0.105

+0.02/°C

±0.170

+0.03/°C