beautypg.com

Atec Agilent-E8364B User Manual

Page 52

background image

48

Table 16. 85050B Calibration Kit

Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)

Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch

Applies to the, E836xB/C analyzers, 85050B (7mm) calibration kit, 85132F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

0.045 to

2 GHz

2 to

18 GHz

Directivity

52

52

Source Match

48

41

Load Match

52

47

Reflection Tracking

±0.003

+0.02/°C

±0.047

+0.02/°C

Transmission Tracking

±0.008

+0.02/°C

±0.034

+0.02/°C