Atec Agilent-E8364B User Manual
Page 70
66
Table 22. 85054B Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E836xB/C analyzers, 85054B (Type-N) calibration kit, 85132F flexible test port cable set with 85130C adapter set, and a full
2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
0.045 to
2 GHz
2 to
18 GHz
Directivity
48
42
Source Match
45
33
Load Match
48
41
Reflection Tracking
±0.001
+0.02/°C
±0.015
+0.02/°C
Transmission Tracking
±0.011
+0.02/°C
±0.083
+0.02/°C