Atec Agilent-E8364B User Manual
Page 114
110
Table 35 (Continued). Test Port Input
Third Order Intercept - Tone spacing from 100 kHz - 5 MHz
Typical
10 MHz to 150
MHz
--
+33 dBm
150 MHz to 300
MHz
--
+34 dBm
300 MHz to 500
MHz
--
+30 dBm
500 MHz to 20
GHz
--
+24 dBm
20 to 40 GHz
--
+18 dBm
40 to 50 GHz
--
+15 dBm
Third Order Intercept - Tone spacing from 5 MHz - 20 MHz
Typical
10 MHz to 500
MHz
--
+20 dBm
500 MHz to 20
GHz
--
+20 dBm
20 to 40 GHz
--
+16 dBm
40 to 50 GHz
--
+15 dBm
Third Order Intercept - Tone spacing from 20 MHz - 50 MHz
Typical
10 MHz to 500
MHz
--
+26 dBm
500 MHz to 20
GHz
--
+26 dBm
20 to 40 GHz
--
+20 dBm
40 to 50 GHz
--
+19 dBm
Trace Noise Magnitude
1 kHz IF bandwidth. Ratio measurement, nominal power at test port.
10 MHz to
45 MHz
b
--
<0.050 dB rms
(typical)
45 MHz to
500 MHz
e
<0.010 dB rms
--
500 MHz to
20 GHz
<0.006 dB rms
--
20 GHz to
40 GHz
<0.006 dB rms
--
40 GHz to
50 GHz
<0.006 dB rms
--