Atec Agilent-E8364B User Manual

Page 34

background image

30

Table 10. 85052B Calibration Kit

Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)

Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch

Applies to the, E836xB/C analyzers, 85052B (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Data
and traces above 20 GHz are not applicable to the E8362C. Also applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

45 MHz to

2 GHz

2 to

20 GHz

20 to

26.5 GHz

Directivity

48

44

44

Source Match

40

31

31

Load Match

48

44

44

Reflection Tracking

±0.003

+0.02/°C

±0.006

+0.02/°C

±0.006

+0.03/°C

Transmission Tracking

±0.017

+0.02/°C

±0.091

+0.02/°C

±0.106

+0.03/°C