Atec Agilent-E8364B User Manual
Page 16
12
Table 4. 85056A Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E8363B/C AND E8364B/C analyzers, 85056A (2.4mm) calibration kit, 85133F flexible test port
cable set, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
45 MHz to
2 GHz
2 to
20 GHz
20 to
40 GHz
40 to
50 GHz
Directivity
42
42
38
36
Source Match
41
38
33
31
Load Match
42
42
37
35
Reflection Tracking
±0.001
+0.02/°C
±0.008
+0.02/°C
±0.020
+0.02/°C
±0.027
+0.03/°C
Transmission Tracking
±0.019
+0.02/°C
±0.053
+0.02/°C
±0.109
+0.02/°C
±0.182
+0.03/°C