beautypg.com

Atec Agilent-E8364B User Manual

Page 64

background image

60

Table 20. 85050D Calibration Kit

Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)

Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch

Applies to the, E836xB/C analyzers, 85050D (7mm) calibration kit, 85132F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

0.045 to

2 GHz

2 to

18 GHz

Directivity

40

40

Source Match

39

35

Load Match

40

37

Reflection Tracking

±0.010

+0.02/°C

±0.100

+0.02/°C

Transmission Tracking

±0.025

+0.02/°C

±0.078

+0.02/°C