Atec Agilent-E8364B User Manual
Page 64
60
Table 20. 85050D Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E836xB/C analyzers, 85050D (7mm) calibration kit, 85132F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
0.045 to
2 GHz
2 to
18 GHz
Directivity
40
40
Source Match
39
35
Load Match
40
37
Reflection Tracking
±0.010
+0.02/°C
±0.100
+0.02/°C
Transmission Tracking
±0.025
+0.02/°C
±0.078
+0.02/°C