Atec Agilent-E8364B User Manual
Page 125
121
Table 40 (Continued). Rear Panel Information
USB Port
One port on front panel and five ports on rear panel. Universal Serial Bus jack,
Type A configuration (4 contacts inline, contact 1 on left); female
Contact 1
Vcc: 4.75 to 5.25 VDC, 500 mA, maximum
Contact 2
-Data
Contact 3
+Data
Contact 4
Ground
LAN
10/100BaseT Ethernet, 8-pin configuration; auto selects between the two data
rates
Line Power
5
Frequency, Voltage
50/60/400 Hz for 100 – 120 V
50/60 Hz for 220 – 240 V
Power supply is auto switching
Max
350 W
1
Measured at -5 dBm test port power.
2
Pulse input connectors are operational only with Option H08 (Pulse Measurement Capability) enabled.
3
Based on deviation from signal reduction equation: Signal Reduction (dB) = 20log
10
(Duty_cycle) =
20log
10
(pulse_width/pulse_repetition_interval). Measured at pulse repetition frequency of 1 MHz.
4
Test port power has to be at a high enough level such that the Drop Cal does not occur. If Drop Cal occurs then the power out of the rear
panel RF connector will drop by about 15 dB.
5
A third-wire ground is required.