beautypg.com

Atec Agilent-E8364B User Manual

Page 19

background image

15

Table 5. 85056D Calibration Kit

Standard Configuration and Standard Power Range

(E8363B/C AND E8364B/C)

Applies to the, E8363B/C AND E8364B/C analyzers, 85056D (2.4mm) calibration kit, 85133F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

45 MHz to

2 GHz

2 to

20 GHz

20 to

40 GHz

40 to

50 GHz

Directivity

42

34

26

26

Source Match

40

30

24

23

Load Match

42

33

25

25

Reflection Tracking

±0.002

+0.02/°C

±0.029

+0.02/°C

±0.079

+0.02/°C

±0.075

+0.03/°C

Transmission Tracking

±0.011

+0.02/°C

±0.121

+0.02/°C

±0.347

+0.02/°C

±0.462

+0.03/°C

This manual is related to the following products: