45

E836xB/C Corrected System Performance with 7mm Connectors

Table 15. 85050B Calibration Kit

Standard Configuration and Standard Power Range

(E836xB/C)

Applies to the, E836xB/C analyzers, 85050B (7mm)

calibration kit, 85132F flexible test port cable set, and a full 2-port

calibration. Also applies to the f

ollowing condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

0.045 to

2 GHz

2 to

18 GHz

Directivity

52

52

Source Match

48

41

Load Match

52

47

Reflection Tracking

±0.003

+0.02/°C

±0.047

+0.02/°C

Transmission Tracking

±0.004

+0.02/°C

±0.032

+0.02/°C

This manual is related to the following products: