Measurement throughput summary – Atec Agilent-E8364B User Manual
Page 127
123
Measurement Throughput Summary
Table 42 Typical Cycle Time
a,b
(ms) for Measurement Completion
Number of Points
201
401
1601
16,001
Start 28 GHz, Stop 30 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
12
19
55
503
2-Port cal
29
44
124
1112
Start 10 MHz, Stop 10 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
86
93
121
583
2-Port cal
179
199
267
1301
Start 10 MHz, Stop 20 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
126
130
153
597
2-Port cal
264
275
335
1321
Start 10 MHz, Stop 40 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
185
190
213
621
2-Port cal
382
401
459
1374
Start 10 MHz, Stop 50 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
210
216
243
643
2-Port cal
436
450
522
1405
Start 10 MHz, Stop 67 GHz, 35 kHz IF bandwidth
Uncorrected
1-Port cal
244
254
300
645
2-Port cal
502
524
591
1423
a
Typical performance.
b
Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with