Measurement Throughput Summary

Table 42 Typical Cycle Time

a,b

(ms) for Measurement Completion

Number of Points

201

401

1601

16,001

Start 28 GHz, Stop 30 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

12

19

55

503

2-Port cal

29

44

124

1112

Start 10 MHz, Stop 10 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

86

93

121

583

2-Port cal

179

199

267

1301

Start 10 MHz, Stop 20 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

126

130

153

597

2-Port cal

264

275

335

1321

Start 10 MHz, Stop 40 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

185

190

213

621

2-Port cal

382

401

459

1374

Start 10 MHz, Stop 50 GHz, 35 kHz IF bandwidth

Uncorrected,

1-port cal

210

216

243

643

2-Port cal

436

450

522

1405

Start 10 MHz, Stop 67 GHz, 35 kHz IF bandwidth

Uncorrected

1-Port cal

244

254

300

645

2-Port cal

502

524

591

1423

a

Typical performance.

b

Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with

DISPLAY:ENABLE OFF. Add 21 ms for display on. Data for one trace (S11) measurement.