beautypg.com

Atec Agilent-E8364B User Manual

Page 25

background image

21

E8363B/C AND E8364B/C Corrected System Performance with 2.92mm Connectors

Table 7. 85056K Calibration Kit

Standard Configuration and Standard Power Range

(E8363B/C AND E8364B/C)

Applies to the, E8363B/C AND E8364B/C analyzers, 85056K (2.92mm) calibration kit, 85133F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

0.045 to

2 GHz

2 to

20 GHz

20 to

40 GHz

Directivity

42

42

40

Source Match

40

40

35

Load Match

42

41

38

Reflection Tracking

±0.018

+0.02/°C

±0.018

+0.02/°C

±0.067

+0.03/°C

Transmission Tracking

±0.011

+0.02/°C

±0.042

+0.02/°C

±0.089

+0.03/°C

This manual is related to the following products: