Atec Agilent-E8364B User Manual
Page 25
21
E8363B/C AND E8364B/C Corrected System Performance with 2.92mm Connectors
Table 7. 85056K Calibration Kit
Standard Configuration and Standard Power Range
(E8363B/C AND E8364B/C)
Applies to the, E8363B/C AND E8364B/C analyzers, 85056K (2.92mm) calibration kit, 85133F flexible test port cable set, and a full 2-port
calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
0.045 to
2 GHz
2 to
20 GHz
20 to
40 GHz
Directivity
42
42
40
Source Match
40
40
35
Load Match
42
41
38
Reflection Tracking
±0.018
+0.02/°C
±0.018
+0.02/°C
±0.067
+0.03/°C
Transmission Tracking
±0.011
+0.02/°C
±0.042
+0.02/°C
±0.089
+0.03/°C