Definitions

All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes
after the instrument has been turned on.

Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected
statistical performance distribution, measurement uncertainties, and changes in performance due to
environmental conditions.

Characteristic (char.): A performance parameter that the product is expected to meet before it leaves the
factory, but that is not verified in the field and is not covered by the product warranty. A characteristic includes
the same guardbands as a specification.

Typical (typ.): Expected performance of an average unit which does not include guardbands. It is not covered
by the product warranty.

Nominal (nom.): A general, descriptive term that does not imply a level of performance. It is not covered by the
product warranty.

Calibration: The process of measuring known standards to characterize a network analyzer's systematic
(repeatable) errors.

Corrected (residual): Indicates performance after

error correction

(calibration). It is determined by the quality of

calibration standards and how well "known" they are, plus system repeatability, stability, and noise.

Uncorrected (raw): Indicates instrument performance without error correction. The uncorrected performance
affects the stability of a

calibration

.

Standard: When referring to the analyzer, this includes no options unless noted otherwise.

⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯

Corrected System Performance

The specifications in this section apply for measurements made with the E836xB/C analyzer with the following
conditions:

ƒ

10 Hz IF bandwidth

ƒ

No averaging applied to data

ƒ

Isolation calibration with an averaging factor of 8