Atec Agilent-E8364B User Manual
Page 111
107
Table 35 (Continued). Test Port Input
Test Port Noise Floor
a,b
Option 080 enabled
d
10 Hz IF Bandwidth
10 MHz to
45 MHz
b
<-77 dBm
(typical)
<-77 dBm
(typical)
<-77 dBm
(typical)
<-77 dBm
(typical)
--
45 MHz to
500 MHz
c
<-88 dBm
<-88 dBm
<-88 dBm
<-88 dBm
--
500 MHz to
2 GHz
<-113 dBm
<-113 dBm
<-113 dBm
<-113 dBm
--
2 GHz to
10 GHz
<-116 dBm
<-116 dBm
<-116 dBm
<-116 dBm
--
10 GHz to
20 GHz
<-118 dBm
<-118 dBm
<-118 dBm
<-118 dBm
--
20 GHz to
40 GHz
<-112 dBm
<-112 dBm
<-112 dBm
<-112 dBm
40 GHz to
50 GHz
<-111 dBm
<-111 dBm
<-111 dBm
<-111 dBm
Option 016
degrades
performance by
2 dB.
1 KHz IF Bandwidth
10 MHz to
45 MHz
b
<-57 dBm
(typical)
<-57 dBm
(typical)
<-57 dBm
(typical)
<-57 dBm
(typical)
--
45 MHz to
500 MHz
c
<-68 dBm
<-68 dBm
<-68 dBm
<-68 dBm
--
500 MHz to
2 GHz
<-93 dBm
<-93 dBm
<-93 dBm
<-93 dBm
--
2 GHz to
10 GHz
<-96 dBm
<-96 dBm
<-96 dBm
<-96 dBm
--
10 GHz to
20 GHz
<-98 dBm
<-98 dBm
<-98 dBm
<-98 dBm
--
20 GHz to
40 GHz
<-92 dBm
<-92 dBm
<-92 dBm
<-92 dBm
40 GHz to
50 GHz
<-91 dBm
<-91 dBm
<-91 dBm
<-91 dBm
Option 016
degrades
performance by
2 dB.