Atec Agilent-E8364B User Manual
Page 113
109
Table 35 (Continued). Test Port Input
Direct Receiver Access Input Noise Floor
a,b
- Option 080 enabled
d
10 Hz IF Bandwidth
10 MHz to 45
MHz
--
<-127 dBm
--
<-127 dBm
--
45 MHz to 500
MHz
c
--
<-126 dBm
--
<-126 dBm
--
500 MHz to 2
GHz
--
<-132 dBm
--
<-132 dBm
--
2 GHz to
10 GHz
--
<-131 dBm
--
<-131 dBm
--
10 GHz to 20
GHz
--
<-133 dBm
--
<-133 dBm
--
20 GHz to 40
GHz
--
<-124 dBm
--
<-124 dBm
40 GHz to 50
GHz
--
<-122 dBm
--
<-122 dBm
Option 016
degrades
performance by
2 dB.
1 KHz IF Bandwidth
10 MHz to 45
MHz
--
<-107 dBm
--
<-107 dBm
--
45 MHz to
500 MHz
c
--
<-106 dBm
--
<-106 dBm
--
500 MHz to 2
GHz
--
<-112 dBm
--
<-112 dBm
--
2 GHz to
10 GHz
--
<-111 dBm
--
<-111 dBm
--
10 GHz to
20 GHz
--
<-113 dBm
--
<-113 dBm
--
20 GHz to
40 GHz
--
<-104 dBm
--
<-104 dBm
40 GHz to
50 GHz
--
<-102 dBm
--
<-102 dBm
Option 016
degrades
performance by
2 dB.
Receiver Compression Level (Measured at Test Ports)
10 MHz to 20
GHz
<0.1 dB at -5 dBm
g
and <0.45 dB at +5 dBm
--
20 GHz to
30 GHz
<0.1 dB at -9.5 dBm
g
and <0.45 dB at 0 dBm
--
30 GHz to
40 GHz
<0.1 dB at -12.5 dBm
g
and <0.45 dB at -3 dBm
--
40 GHz to
50 GHz
<0.1 dB at -12.5 dBm
g
and <0.45 dB at -3 dBm
--
System Compression Level
maximum output power
See dynamic
accuracy table