Table 12. 85052C Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E836xB/C analyzers, 85052C (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Data
and traces above 20 GHz are not applicable to the E8362C. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
45 MHz to
2 GHz
2 to
20 GHz
20 to
26.5 GHz
Directivity
48
50
50
Source Match
40
50
50
Load Match
48
50
50
Reflection Tracking
±0.003
+0.02/°C
±0.000
+0.02/°C
±0.000
+0.03/°C
Transmission Tracking
±0.017
+0.02/°C
±0.016
+0.02/°C
±0.019
+0.03/°C