69

Table 23. 85054D Calibration Kit

Standard Configuration and Standard Power Range

(E836xB/C)

Applies to the, E836xB/C analyzers, 85054D (Type-N) calibration kit, 85132F flexible test port cable set with 85130C adapter set, and a full
2-port calibration. Also applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

0.045 to

2 GHz

2 to

18 GHz

Directivity

40

34

Source Match

39

29

Load Match

40

34

Reflection Tracking

±0.003

+0.02/°C

±0.027

+0.02/°C

Transmission Tracking

±0.013

+0.02/°C

±0.136

+0.02/°C