beautypg.com

Atec Agilent-E8364B User Manual

Page 55

background image

51

Table 17. 85050C Calibration Kit

Standard Configuration and Standard Power Range

(E836xB/C)

Applies to the, E836xB/C analyzers, 85050C (7mm) calibration kit, 85132F flexible test port cable set, and a full 2-port calibration. Also
applies to the following condition:

Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature

Specification (dB)

Description

0.045 to

2 GHz

2 to

18 GHz

Directivity

52

60

Source Match

48

60

Load Match

52

60

Reflection Tracking

±0.003

+0.02/°C

±0.000

+0.02/°C

Transmission Tracking

±0.004

+0.02/°C

±0.004

+0.02/°C

This manual is related to the following products: