Atec Agilent-E8364B User Manual
Page 109
105
Table 34 (Continued). Test Port Output
1 MHz offset from center frequency, nominal power at test port - Option 080 enabled
Typical:
10 MHz to 10
GHz
--
--
--
--
-103 dBc
10 GHz to 20
GHz
--
--
--
--
-97 dBc
20 GHz to 50
GHz
--
--
--
--
-85 dBc
Harmonics (2nd or 3rd)
-23 dBc typical,
in power range
0
Non-Harmonic Spurious (at Nominal Output Power)
10 MHz to 45
MHz
--
--
--
--
-50 dBc typical,
for offset
frequency > 1
kHz
45 MHz to 20
GHz
--
--
--
--
-50 dBc typical,
for offset
frequency > 1
kHz
20 GHz to 40
GHz
--
--
--
--
-30 dBc typical,
for offset
frequency > 1
kHz
40 GHz to 50
GHz
--
--
--
--
-30 dBc typical,
for offset
frequency > 1
kHz
a
Test port output is a specification when the source is set to Port 1, and a characteristic when the source is set to Port 2.
b
Typical performance.
c
Preset power.
d
Power Level Linearity is a specification when the source is set to Port 1, and a typical when the source is set to Port 2.
e
Test port power is specified into nominal 50 ohms.
f
Power to which the source can be set and phase lock is assured.
g
+/-1.5 dB for power <= -23 dBm.
h
E8362B/C only: Option H11 decreases maximum power level by 1 dB.
i
E8362B/C only: Option H11 decreases maximum power level by 2 dB.
j
E8362B/C only: Option H11 decreases power level by 1 dB.
k
E8362B/C only: Option H11 decreases power level by 2 dB.
l
Assumes no variation in time.
m
Assumes no variation in temperature.