Atec Agilent-E8364B User Manual
Page 82
78
Table 26. R11644A Calibration Kit
Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081)
Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offset Mode, and Reference Channel
Transfer Switch
Applies to the, E8363B/C AND E8364B/C analyzers, R11644A (WR-28) calibration kit, 85133F flexible test port cable set with the R281A
and R281B launch sets, and a full 2-port calibration. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB)
Description
26.5 to
40 GHz
Directivity
50
Source Match
50
Load Match
50
Reflection Tracking
±0.000
+0.03/°C
Transmission Tracking
±0.019
+0.03/°C
This manual is related to the following products: