124
Table 43. Cycle Time vs IF Bandwidth
a
Applies to the Preset condition
(
201 points, correction off) except for the following changes:
CF = 28 GHz
Span = 100 MHz
Display off (add 21 ms for display on)
IF
Bandwidth
(Hz)
Cycle Time
(ms)
b
Cycle Time
(ms)
Option 080
enabled
40,000
11
100
35,000
12
101
30,000
13
102
20,000
16
106
10,000
30
127
7000
38
138
5000
50
152
3000
74
182
1000
274
326
300
694
782
100
1905
2054
30
6091
6355
10
17916
18372
a
Typical performance.
b
Cycle time includes sweep and retrace time.
Table 44. Cycle Time vs Number of Points
a
Applies to the Preset condition (35 kHz IF bandwidth, correction off) except for the following changes:
CF = 28 GHz
Span = 100 MHz
Display off (add 21 ms for display on)
Number
of
Points
Cycle Time
(ms)
b
3
6
11
6
51
7
101
9
201
12
401
18
801
30
1601
55
16,001
497
a
Typical performance.
b
Cycle time includes sweep and retrace time.