beautypg.com

Atec Agilent-E8364B User Manual

Page 128

background image

124

Table 43. Cycle Time vs IF Bandwidth

a

Applies to the Preset condition

(

201 points, correction off) except for the following changes:

ƒ

CF = 28 GHz

ƒ

Span = 100 MHz

ƒ

Display off (add 21 ms for display on)

IF
Bandwidth
(Hz)

Cycle Time
(ms)

b

Cycle Time
(ms)

Option 080
enabled

40,000

11

100

35,000

12

101

30,000

13

102

20,000

16

106

10,000

30

127

7000

38

138

5000

50

152

3000

74

182

1000

274

326

300

694

782

100

1905

2054

30

6091

6355

10

17916

18372

a

Typical performance.

b

Cycle time includes sweep and retrace time.

Table 44. Cycle Time vs Number of Points

a

Applies to the Preset condition (35 kHz IF bandwidth, correction off) except for the following changes:

ƒ

CF = 28 GHz

ƒ

Span = 100 MHz

ƒ

Display off (add 21 ms for display on)

Number
of

Points

Cycle Time
(ms)

b

3

6

11

6

51

7

101

9

201

12

401

18

801

30

1601

55

16,001

497

a

Typical performance.

b

Cycle time includes sweep and retrace time.

This manual is related to the following products: