Atec Agilent-E8364B User Manual
Page 110
106
Table 35: Test Port Input
Description
Specification
Supplemental
Standard
Opt 014
Opt UNL
Opt 014/UNL
Test Port Noise Floor
a
10 Hz IF Bandwidth
10 MHz to
45 MHz
b
<-77 dBm
(typical)
<-77 dBm
(typical)
<-77 dBm
(typical)
<-77 dBm
(typical)
--
45 MHz to
500 MHz
c
<-89 dBm
<-89 dBm
<-89 dBm
<-89 dBm
--
500 MHz to
2 GHz
<-114 dBm
<-114 dBm
<-114 dBm
<-114 dBm
--
2 GHz to
10 GHz
<-117 dBm
<-117 dBm
<-117 dBm
<-117 dBm
--
10 GHz to
20 GHz
<-120 dBm
<-119 dBm
<-120 dBm
<-119 dBm
--
20 GHz to
40 GHz
<-114 dBm
<-113 dBm
<-114 dBm
<-113 dBm
40 GHz to
50 GHz
<-114 dBm
<-112 dBm
<-114 dBm
<-112 dBm
Option 016
degrades
performance by
2 dB.
1 KHz IF Bandwidth
10 MHz to
45 MHz
b
<-57 dBm
(typical)
<-57 dBm
(typical)
<-57 dBm
(typical)
<-57 dBm
(typical)
--
45 MHz to
500 MHz
c
<-69 dBm
<-69 dBm
<-69 dBm
<-69 dBm
--
500 MHz to
2 GHz
<-94 dBm
<-94 dBm
<-94 dBm
<-94 dBm
--
2 GHz to
10 GHz
<-97 dBm
<-97 dBm
<-97 dBm
<-97 dBm
--
10 GHz to
20 GHz
<-100 dBm
<-99 dBm
<-100 dBm
<-99 dBm
--
20 GHz to
40 GHz
<-94 dBm
<-93 dBm
<-94 dBm
<-93 dBm
40 GHz to
50 GHz
<-94 dBm
<-92 dBm
<-94 dBm
<-92 dBm
Option 016
degrades
performance by
2 dB.