beautypg.com

Atec Agilent-E8364B User Manual

Page 110

background image

106

Table 35: Test Port Input

Description

Specification

Supplemental

Standard

Opt 014

Opt UNL

Opt 014/UNL

Test Port Noise Floor

a

10 Hz IF Bandwidth

10 MHz to

45 MHz

b

<-77 dBm
(typical)

<-77 dBm
(typical)

<-77 dBm
(typical)

<-77 dBm
(typical)

--

45 MHz to

500 MHz

c

<-89 dBm

<-89 dBm

<-89 dBm

<-89 dBm

--

500 MHz to

2 GHz

<-114 dBm

<-114 dBm

<-114 dBm

<-114 dBm

--

2 GHz to

10 GHz

<-117 dBm

<-117 dBm

<-117 dBm

<-117 dBm

--

10 GHz to

20 GHz

<-120 dBm

<-119 dBm

<-120 dBm

<-119 dBm

--

20 GHz to

40 GHz

<-114 dBm

<-113 dBm

<-114 dBm

<-113 dBm

40 GHz to

50 GHz

<-114 dBm

<-112 dBm

<-114 dBm

<-112 dBm

Option 016
degrades
performance by
2 dB.

1 KHz IF Bandwidth

10 MHz to

45 MHz

b

<-57 dBm
(typical)

<-57 dBm
(typical)

<-57 dBm
(typical)

<-57 dBm
(typical)

--

45 MHz to

500 MHz

c

<-69 dBm

<-69 dBm

<-69 dBm

<-69 dBm

--

500 MHz to

2 GHz

<-94 dBm

<-94 dBm

<-94 dBm

<-94 dBm

--

2 GHz to

10 GHz

<-97 dBm

<-97 dBm

<-97 dBm

<-97 dBm

--

10 GHz to

20 GHz

<-100 dBm

<-99 dBm

<-100 dBm

<-99 dBm

--

20 GHz to

40 GHz

<-94 dBm

<-93 dBm

<-94 dBm

<-93 dBm

40 GHz to

50 GHz

<-94 dBm

<-92 dBm

<-94 dBm

<-92 dBm

Option 016
degrades
performance by
2 dB.

This manual is related to the following products: