beautypg.com

Front panel jumpers – Atec Agilent-E8364B User Manual

Page 130

background image

126

Specifications: Front-Panel Jumpers

⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯

Models E8362B/C, E8363B/C, and E8364B/C Option 014

⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯

NOTE: The standard E8362/3/4A/B has no front-panel jumpers.

⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯

Table 47: Measurement Receiver Inputs (Rcvr A In, Rcvr B In)

Description

Specification

Supplemental Information

Maximum Input Level

E8362B/C: Typical

45 MHz to 500 MHz

--

-15 dBm, typical

500 MHz to 2 GHz

--

-11 dBm, typical

2 GHz to 10 GHz

--

-11 dBm, typical

10 GHz to 20 GHz

--

-11 dBm, typical

E8363B/C:

45 MHz to 500 MHz

--

-14 dBm, typical

500 MHz to 2 GHz

--

-10 dBm, typical

2 GHz to 10 GHz

--

-10 dBm, typical

10 GHz to 20 GHz

--

-10 dBm, typical

20 GHZ to 30 GHz

--

-14.5 dBm, typical

30 GHZ to 40 GHz

--

-16.5 dBm, typical

E8364B/C:

45 MHz to 500 MHz

--

- 14 dBm, typical

500 MHz to 2 GHz

--

- 10 dBm, typical

2 GHz to 10 GHz

--

- 10 dBm, typical

10 GHz to 20 GHz

--

- 10 dBm, typical

20 GHZ to 30 GHz

--

- 14.5 dBm, typical

30 GHZ to 40 GHz

--

- 16.5 dBm, typical

40 GHZ to 45 GHz

--

- 16 dBm, typical

45 GHZ to 50 GHz

--

- 15 dBm, typical

Noise Floor

E8362B/C:

10 Hz IF Bandwidth

45 MHz to 500 MHz

< -109 dBm

--

500 MHz to 2 GHz

< -130 dBm

--

2 GHz to 10 GHz

< -133 dBm

--

10 GHz to 20 GHz

< -135 dBm

--

1 kHz IF Bandwidth

45 MHz to 500 MHz

< -89 dBm

--

500 MHz to 2 GHz

< -110 dBm

--

2 GHz to 10 GHz

< -113 dBm

--

10 GHz to 20 GHz

< -115 dBm

--

This manual is related to the following products: