Cleaning Information

Which Cleaning Procedure should I use ?

98

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

Preferred Procedure

Use the following procedure on most occasions.

1 Clean the connector by rubbing a new, dry cotton swab over the

surface using a small circular movement.

2 Blow away any remaining lint with compressed air.

Procedure for Stubborn Dirt

Use this procedure when there is greasy dirt on the connector:

1 Moisten a new cotton swab with isopropyl alcohol.

2 Clean the connector by rubbing the cotton swab over the surface

using a small circular movement.

3 Take a new, dry soft tissue and remove the alcohol, dissolved

sediment and dust, by rubbing gently over the surface using a small

circular movement.

4 Blow away any remaining lint with compressed air.

An Alternative Procedure

A better, more gentle, but more expensive cleaning procedure is to use

an ultrasonic bath with isopropyl alcohol.

1 Hold the tip of the connector in the bath for at least three minutes.

2 Take a new, dry soft tissue and remove the alcohol, dissolved

sediment and dust, by rubbing gently over the surface using a small

circular movement.

3 Blow away any remaining lint with compressed air.

How to clean connector adapters

C A U T I O N

Some adapters have an anti-reflection coating on the back to reduce

back reflection. This coating is extremely sensitive to solvents and

mechanical abrasion. Extra care is needed when cleaning these

adapters.

Preferred Procedure

Use the following procedure on most occasions.

1 Clean the adapter by rubbing a new, dry cotton swab over the

surface using a small circular movement.

2 Blow away any remaining lint with compressed air.

Procedure for Stubborn Dirt

Use this procedure when there is greasy dirt on the adapter:

1 Moisten a new cotton swab with isopropyl alcohol.

2 Clean the adapter by rubbing the cotton swab over the surface using

a small circular movement.