Cleaning Information

What do I need for proper cleaning?

92

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

We suggest these protective coverings should be kept on the

equipment at all times, except when your optical device is in use. Be

careful when replacing dust caps after use. Do not press the bottom of

the cap onto the fiber too hard, as any dust in the cap can scratch or

pollute your fiber surface.

If you need further dust caps, please contact your nearest Agilent

Technologies sales office.

Isopropyl alcohol

This solvent is usually available from any local pharmaceutical

supplier or chemist's shop.

If you use isopropyl alcohol to clean your optical device, do not

immediately dry the surface with compressed air (except when you

are cleaning very sensitive optical devices). This is because the dust

and the dirt is solved and will leave behind filmy deposits after the

alcohol is evaporated. You should therefore first remove the alcohol

and the dust with a soft tissue, and then use compressed air to blow

away any remaining filaments.

If possible avoid using denatured alcohol containing additives.

Instead, apply alcohol used for medical purposes.

Never drink this alcohol, as it may seriously damage your health.

Do not use any other solvents, as some may damage plastic materials

and claddings. Acetone, for example, will dissolve the epoxy used with

fiber optic connectors. To avoid damage, only use isopropyl alcohol.

Cotton swabs

We recommend that you use swabs such as Q-tips or other cotton

swabs normally available from local distributors of medical and

hygiene products (for example, a supermarket or a chemist's shop).

You may be able to obtain various sizes of swab. If this is the case,

select the smallest size for your smallest devices.

Ensure that you use natural cotton swabs. Foam swabs will often leave

behind filmy deposits after cleaning.

Use care when cleaning, and avoid pressing too hard onto your optical

device with the swab. Too much pressure may scratch the surface, and

could cause your device to become misaligned. It is advisable to rub

gently over the surface using only a small circular movement.

Swabs should be used straight out of the packet, and never used twice.

This is because dust and dirt in the atmosphere, or from a first

cleaning, may collect on your swab and scratch the surface of your

optical device.