Performance Tests

Performance Tests

42

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

The thermal characteristics of the sphere do not change from device to

device, or during the product lifetime. If a head without the sphere

passes the linearity test within the specifications given on the Test

Record Sheet, then the whole system - sphere and head - is within the

product specification.

The head and the integrating sphere of the 81628B are adjusted and

serialized together, therefore the specifications of the 81628B are only

valid for that given combination of head and integrating sphere.

The disconnecting and connecting of the integrating sphere may

change the performance verification data slightly.