Performance Tests

Performance Tests

56

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

b

Set the wavelength [

λ] to the wavelength of the RTL source.

c

Set [Pwr unit] to .

7 At the Return Loss Module:

a

Set the averaging time [AvTime] to 1s.

b

Enter the Return Loss Reference value RLref of the 81610CC
reference cable for this wavelength.

c

Press [RefCal to calibrate the Return Loss module at reference
condition.

8 At the Power Meter press [Disp -> Ref]. The Power Meter should now

read 0.0 dB.

9 Disconnect the 81610CC Reference Cable from the setup and

connect a 81113SC user cable to the Return Loss Module and leave

the other end open as shown in Figure 8.

Figure 14

Return Loss Measurement Setup

10 At the Return Loss Module, press [TermCal] in order to calibrate

the Return Loss Module at termination condition.

N O T E

Do not wrap the cable. The termination condition is realized with the

open end of the patchcord.

11 Now connect the open end to the optical head.

12 Enter the Power reading [P] on the optical head as [FPDelta] on the

Return Loss Meter. Make sure you enter the sign of this value

correctly.

13 Note the Return Loss reading as the Return Loss result into the test

record.

81113SC

Mainframe 8163A/B w/ 8161xA RTL
and 81618A Interface Module

8163A/B

8161xA

in

out

81618A

Optical Head 8162xA

DUT