Performance Tests

Performance Tests

66

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

N O T E

The nonlinearity of the 81622B is not regularly tested up to the

specified power level of +27 dBm. Instead, limited testing up to

+9 dBm is used to test the electronic circuitry of the 81622B. Above

+9 dBm, the largest contribution to nonlinearity comes from the

absorbing glass filter (which is tested on a sample basis) that does not

change its nonlinearity with time. This way it is guaranteed by design

that testing the 81622B up to +9 dBm ensures specification compliance

up to +27 dBm.