background image

What do I need for proper cleaning?

Cleaning Information

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

93

Soft tissues

These are available from most stores and distributors of medical and

hygiene products such as supermarkets or chemists' shops.

We recommend that you do not use normal cotton tissues, but multi-

layered soft tissues made from non-recycled cellulose. Cellulose

tissues are very absorbent and softer. Consequently, they will not

scratch the surface of your device over time.

Use care when cleaning, and avoid pressing on your optical device

with the tissue. Pressing too hard may lead to scratches on the surface

or misalignment of your device. Just rub gently over the surface using

a small circular movement.

Use only clean, fresh soft tissues and never apply them twice. Any dust

and dirt from the air which collects on your tissue, or which has

gathered after initial cleaning, may scratch and pollute your optical

device.

Pipe cleaner

Pipe cleaners can be purchased from tobacconists, and come in

various shapes and sizes.The most suitable one to select for cleaning

purposes has soft bristles, which will not produces scratches.

There are many different kinds of pipe cleaner available from tobacco

nists.

The best way to use a pipe cleaner is to push it in and out of the device

opening (for example, when cleaning an interface). While you are

cleaning, you should slowly rotate the pipe cleaner.

Only use pipe cleaners on connector interfaces or on feed through

adapters. Do not use them on optical head adapters, as the center of a

pipe cleaner is hard metal and can damage the bottom of the adapter.

Your pipe cleaner should be new when you use it. If it has collected

any dust or dirt, this can scratch or contaminate your device.

The tip and center of the pipe cleaner are made of metal. Avoid

accidentally pressing these metal parts against the inside of the device,

as this can cause scratches.

Compressed air

Compressed air can be purchased from any laboratory supplier.

It is essential that your compressed air is free of dust, water and oil.

Only use clean, dry air. If not, this can lead to filmy deposits or

scratches on the surface of your connector. This will reduce the

performance of your transmission system.