Performance Tests

Performance Tests

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

77

Performance Test for the Agilent 81626B

Page 3 of 3

Model Agilent 81626B Optical Head

Report No.________

Date ________________

Test
No.

Test Description

Min.

Spec.

Result

Max.

Spec.

Measurement

Uncertainty

I

Accuracy Test

[

µW]

measured at _______ nm
(1310nm) Output Power

9.64

µW

_______

10.36

µW

measured at _______ nm
(1550nm) Output Power

9.64

µW

_______

10.36

µW

II

Linearity Test

For Calculations you may want to use the appropriate sheet

Range

P

DUT

[dBm]

P

DUT

[dBm]

Loss [%]

+20

+9

________

_______

<± 1.29%

[1]

+10

+9

________

_______

<± 0.93 %

+10

+3

________

_______

<± 0.93 %

0

+3

________

_______

<± 0.93 %

0

-7

________

_______

<± 0.93 %

-10

-7

________

_______

<± 0.93 %

-10

-17

________

_______

<± 0.93 %

-20

-17

________

_______

<± 0.93 %

-20*

-17*

Reference

0.0

<± 0.93 %

-20

-27

________

_______

<± 0.95 %

-30

-27

________

_______

<± 0.95 %

-30

-37

________

_______

<± 1.18 %

-40

-37

________

_______

<± 0.18 %

-40

-47

________

_______

<± 3.43 %

III

Noise Test

[pW]

_______

< 500 pW

IV

Return Loss Test

[dB]

60 dB <

_______

[1] The 20 dB range measurement is only performed to check the functionality of the current booster. The published spec-
ification does not apply to operation in non-automode as used here to check the functionality.