Performance Tests

Performance Tests

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

49

36 Decrease the attenuation of Atty2 to 10 dB in order to be on the

upper limit of the -50 dBm range at the reference power meter.

-20 dBm Range (equivalent to +10 dBm range for
81628B)

37 Disable Atty1.

38 Zero both Power Meters. On the 8163A/B with two installed power

meters, press [Menu], move to <Zero all>, and press Enter.

39 Enable Atty1.

40 Switch one range up to the -10 dBm (20 dBm for the 81628B) range

and note the power readings (#9).

41 Switch down to the previous (-20 dBm, 10 dBm for the 81628B) and

note the power readings again (#10).

42 Increase the attenuation of Atty1 by 10 dB and note the results in

the test record (#11).

43 On the DUT switch one range down to the -30 dBm (0 dBm for the

81628B) range.

44 If necessary, adjust the attenuation of Atty1 in order to be on the

upper limit of the range (i.e. -x7.y dBm).

45 Decrease the attenuation of Atty2 by 10 dB in order to be on the

upper limit of the -50 dBm range.

-30 dBm Range (equivalent to 0 dBm range for 81628B)

46 Disable Atty1.

47 Zero both power meters. On the 8163A/B with two installed power

meters, press [Menu], move to , and press Enter.

48 Enable Atty1.

49 Switch one range up to the -20 dBm range (10 dBm for the 81628B)

and note the power readings (#12).

50 Switch down to the previous range (-30 dBm, 0 dBm for the 81628B)

and note the values again (#13).

51 Increase the attenuation of Atty1 by 10 dB and note the results in

the test record (#14).

52 On the DUT switch one range down to the -40 dBm (-10 for 81628B)

range.