Which Cleaning Procedure should I use ?

Cleaning Information

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

103

3 Blow away any remaining lint with compressed air.

Procedure for Stubborn Dirt

Use this procedure when there is greasy dirt on the interface, and

using the preferred procedure is not sufficient. Using isopropyl

alcohol should be your last choice for recessed lens interfaces because

of the difficulty of cleaning out any dirt that is washed to the edge of

the interface:

1 Moisten a new cotton swab with isopropyl alcohol.

2 Clean the interface by rubbing the cotton swab over the surface

using a small circular movement.

3 Take a new, dry soft tissue and remove the alcohol, dissolved

sediment and dust, by rubbing gently over the surface using a small

circular movement.

4 Blow away any remaining lint with compressed air.

How to clean optical devices which are
sensitive to mechanical stress and pressure

Some optical devices, such as the Agilent 81000BR Reference

Reflector, which has a gold plated surface, are very sensitive to

mechanical stress or pressure. Do not use cotton swabs, soft tissues or

other mechanical cleaning tools, as these can scratch or destroy the

surface.

Preferred Procedure

Use the following procedure on most occasions.

1 Blow away any dust or dirt with compressed air.

Procedure for Stubborn Dirt

To clean devices that are extremely sensitive to mechanical stress or

pressure you can also use an optical clean polymer film. This

procedure is time-consuming, but you avoid scratching or destroying

the surface.

1 Put the film on the surface and wait at least 30 minutes to make

sure that the film has had enough time to dry.

2 Remove the film and any dirt with special adhesive tapes.

Alternative Procedure

For these types of optical devices you can often use an ultrasonic bath

with isopropyl alcohol. Only use the ultrasonic bath if you are sure

that it won't cause any damage any part of the device.

1 Put the device into the bath for at least three minutes.

2 Blow away any remaining liquid with compressed air.