Performance Tests

Performance Tests

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

57

Relative Uncertainty due to Polarization
(Optional Test)

N O T E

The performance test "Relative Uncertainty due to Polarization" is

optional, since the polarization is given with the production of the unit

by mechanical and optical cavities and is unchanged by normal use of

the sensor module.

Refer to Figure 15 for a setup to verify the relative uncertainty due to

polarization of the sensor module.

Generally, during this measurement procedure the tunable laser

source is swept through a predefined wavelength range. After every

wavelength step, a single PDL -measurement is made, where the

polarization controller generates all the different polarization states.

The highest PDL value is taken as the "relative uncertainty due to

polarization". The low output power path of the coupler is used to

monitor the power stability of the setup.