Atec Agilent-81618A User Manual
Page 30
![background image](/manuals/317944/30/background.png)
Specifications
Optical Head Specifications
30
Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition
Notes:
Agilent 81622B
Agilent 81626B
1
Reference Conditions:
• Power level 10
µW (-20 dBm), continuous wave (CW)
• Parallel beam, 3 mm spot diameter on the center of the detector (center of
sphere input for 81628B)
• Ambient temperature 23
°C ± 5 °C
• On day of calibration (add
± 0.3% for aging over one year; add ± 0.6% over
two years)
• Spectral width of source < 10 nm (FWHM)
Wavelength setting at power meter must correspond to source wavelength
±
0.4 nm
2
Operating Conditions:
• Parallel beam, 3mm spot diameter on the center of the detector or
connectorized fiber with NA
≤ 0.2 (straight connector)
• For NA > 0.2, add 1%
• Within one year after calibration, add 0.3% for second year
• Operating temperature range as specified; humidity: non-condensing
3
All states of polarization at constant wavelength (1550 nm ± 30 nm) and
constant power, straight connector, T = 23°C ± 5°. For angled connector
(8°) add 0.01 dB typ.
4
Conditions:
Wavelength 1550 nm ± 30 nm, fixed state of polarization, constant power,
Temperature 23°C ± 5°C,
Linewidth of source
≥ 100 MHz, angled connector 8°.
5
At constant temperature (
∆T = ± 1°C), zeroing required.
6
Averaging time 1s, T = 23°C ± 5°C, observation time 300 s. Wavelength range
1200 - 1630 nm.
7
Conditions
• Wavelengths 1550 nm
± 30 nm.
Standard single-mode fiber, angled connector min 8
°.
8
For input power > 2 mW, add ± 0.004 dB / mW
9
30°C for > +20 dBm input power
10
For input power >+10 mW add:
typ. ± 0.0016 dB / mW without Agilent 81000AF, or add
± 0.0008 dB / mW with Agilent 81000AF (direct coupled)
In the case of a negative power change >50 dB allow an addition recovery
time of 3 minutes.