Cleaning Information

Which Cleaning Procedure should I use ?

102

Agilent 81618A/9A and Agilent 81622B/3B/4B/6B/7B/8B User’s Guide, Second Edition

W A R N I N G

Never look into an optical output, because this can seriously damage

your eyesight.

To assess the projection of the emitted light beam you can use an

infrared sensor card. Hold the card approximately 5 cm from the

interface. The invisible emitted light is projected onto the card and

becomes visible as a small circular spot.

Preferred Procedure

Use the following procedure on most occasions.

1 Clean the interface by rubbing a new, dry cotton swab over the

surface using a small circular movement.

2 Blow away any remaining lint with compressed air.

Procedure for Stubborn Dirt

Use this procedure when there is greasy dirt on the interface:

1 Moisten a new cotton swab with isopropyl alcohol.

2 Clean the interface by rubbing the cotton swab over the surface

using a small circular movement.

3 Take a new, dry soft tissue and remove the alcohol, dissolved

sediment and dust, by rubbing gently over the surface using a small

circular movement.

4 Blow away any remaining lint with compressed air.

How to clean instruments with a recessed
lens interface

W A R N I N G

For instruments with a deeply recessed lens interface (for example the

Agilent 81633A and 81634A Power Sensors) do NOT follow this

procedure. Alcohol and compressed air could damage your lens even

further.

Keep your dust and shutter caps on when your instrument is not in

use. This should prevent it from getting too dirty. If you must clean

such instruments, please refer the instrument to the skilled personnel

of Agilent’s service team.

Preferred Procedure

Use the following procedure on most occasions.

1 Blow away any dust or dirt with compressed air.

If this is not sufficient, then

2 Clean the interface by rubbing a new, dry cotton swab over the

surface using a small circular movement.